Electrospray Charge States of Peptides: How to Read ESI Spectra, Convert m/z to Mass, and Check a Kisspeptin-10 Example

Electrospray charge states of peptides explained with a kisspeptin-10 example: how to find z, convert m/z to mass, and avoid common ESI mistakes.

Electrospray charge states of peptides explained: in electrospray ionization a peptide picks up several protons and appears as a series of ions with different charges so one peptide gives several peaks at different m/z values, the neutral mass is recovered with M = z times (m/z minus 1.00728) for positive-mode protonated ions, two adjacent peaks in a charge series let you work out the charge z without prior knowledge using z = (m2 minus 1.00728) divided by (m1 minus m2) where m1 is the higher m/z and m2 is the lower m/z rounded to the nearest whole number, kisspeptin-10 YNWNSFGLRF-NH2 has a calculated monoisotopic mass of 1301.6305 Da and would be expected mainly at 1+ and 2+ because it has few protonation sites with one arginine and the N-terminal amine, the calculated ions are [M+H]+ at 1302.638, [M+2H]2+ at 651.823, and [M+3H]3+ at 434.884, common mistakes are reading m/z as mass, overlooking sodium and potassium adducts, using the wrong mass definition, neglecting in-source fragments, and treating a single spectrum as purity, BioAura supplies research peptides with batch-specific documentation including a certificate of analysis with identity confirmed by mass spectrometry and purity verified by HPLC at greater than 99%

A research guide to multiply charged peptide ions in electrospray mass spectrometry, with the arithmetic worked through for a ten-residue peptide.

Electrospray charge states are the different numbers of protons a peptide carries in an ESI mass spectrum, so one peptide appears as several peaks at different m/z values. The neutral mass is recovered with M = z x (m/z - 1.00728). This guide works through the calculation for kisspeptin-10, for laboratory researchers only.

Last reviewed: October 2026

Electrospray ionization is the standard way peptide identity is confirmed by mass, but its spectra show mass divided by charge, which confuses newcomers. This article explains what charge states are, how to find the charge and the mass by hand, how a short peptide like kisspeptin-10 behaves, and which mistakes to avoid. It is written strictly for a laboratory research context and describes analytical principles, not any use in people.

Electrospray charge states at a glance

ItemDetail
TechniqueElectrospray ionization (ESI) mass spectrometry [1,2]
ObservationSeries of peaks differing by one charge [1]
Mass formulaM = z x (m/z - 1.00728) for [M+zH]z+ ions
Example peptideKisspeptin-10, YNWNSFGLRF-NH2, 1301.6305 Da (calculated, monoisotopic)
Expected ionsMainly 1+ and 2+ (calculated)
BioAura productKisspeptin-10
Related pagesPeptide mass spectrometry

What are electrospray charge states?

In electrospray ionization (ESI), a peptide picks up several protons and appears as a series of ions with different charges, so one peptide gives several peaks at different m/z values. These are called charge states.

Fenn and colleagues, in the 1989 Science paper that helped establish the technique, described the distinguishing feature of electrospray spectra for large molecules as coherent sequences of peaks whose component ions are multiply charged, with the ions of each peak differing by one charge from those of adjacent neighbors [1]. Banerjee and Mazumdar review the origin and implications of the multiple charge states seen in ESI mass spectra [5]. For a peptide the practical result is that the spectrum does not show the mass directly. It shows mass divided by charge, and a short calculation is needed to recover the mass.

How do you calculate peptide mass from m/z and charge?

The neutral mass is M = z x (m/z - 1.00728) for positive-mode protonated ions, where z is the charge and 1.00728 Da is the mass of a proton. You need to know z first.

Two adjacent peaks in a charge series let you work out z without any prior knowledge. If the lower-m/z peak has charge z+1 and the higher-m/z peak has charge z, then z = (m2 - 1.00728) / (m1 - m2), where m1 is the higher m/z and m2 is the lower m/z, rounded to the nearest whole number. Once z is known for one peak, the neutral mass follows from the first formula. In practice, instrument software performs this deconvolution automatically, but understanding the arithmetic makes it possible to check a result by hand. Cech and Enke review how analyte structure, instrument parameters, and solution composition affect ESI response [3], which explains why the same peptide can give different charge distributions on different instruments.

What charge states does kisspeptin-10 show?

Kisspeptin-10 (YNWNSFGLRF-NH2) has a calculated monoisotopic mass of 1301.6305 Da, and it would be expected mainly at 1+ and 2+, with small contributions from higher states. The reason is that it has few protonation sites.

The sequence is Tyr-Asn-Trp-Asn-Ser-Phe-Gly-Leu-Arg-Phe-NH2, with a C-terminal amide. Its only strongly basic side chain is arginine, and the N-terminal amine adds a second site, so the number of protons it can readily carry is small. The table lists the calculated ions. Values are computed from the sequence and are not measured data.

IonCalculated m/z (monoisotopic)How to get back to the neutral mass
[M+H]+1302.6381302.638 - 1.00728 = 1301.631
[M+2H]2+651.8232 x (651.823 - 1.00728) = 1301.631
[M+3H]3+434.8843 x (434.884 - 1.00728) = 1301.631

Check that the two-charge example works with the adjacent-peak formula: if a 2+ peak appears at 651.823 and a 3+ peak at 434.884, then z for the 2+ peak = (434.884 - 1.00728) / (651.823 - 434.884) = 2.00, which confirms the assignment. BioAura supplies kisspeptin-10 as a research peptide, and our kisspeptin-10 research overview covers the molecule. The C-terminal amidation article explains why the amide changes the mass by -0.984 Da compared with the free acid.

What determines how many charges a peptide carries?

The number of basic sites (Arg, Lys, His, and the N-terminus), the peptide length, the solvent pH, and the instrument conditions determine the charge distribution. Longer peptides with more basic residues reach higher charge states.

Kebarle and Verkerk describe how charged droplets evolve through solvent evaporation to the ions that enter the gas phase, a process that sets how the charge ends up distributed [2]. Cech and Enke discuss analyte chargeability and surface activity in relation to ESI response [3]. Two consequences for bench work follow:

  1. Acidic mobile phases favour protonation. Formic acid is commonly used for this reason, and our TFA vs formic acid article discusses the trade-off with chromatographic performance.
  2. Short peptides show low charge states. A decapeptide with one arginine will look very different from a 39-residue peptide with several basic residues, which can show a wider series of charge states.

How does nanoelectrospray change the picture?

Nanoelectrospray uses a much smaller emitter and a very low flow rate, which gives small droplets and good tolerance for salt. It is widely used when sample is limited.

Wilm and Mann described the nanoelectrospray source as having a spraying orifice of 1-2 microns, a flow rate of roughly 20 nL per minute, small droplets, and no solvent pumps, and reported that solutions with up to 0.1 M salt could be sprayed [4]. That tolerance and sensitivity made it useful for peptide mixtures, and the same paper demonstrated tandem mass spectrometric investigation of an unseparated peptide mixture [4]. For routine identity checks of a purified peptide, conventional electrospray coupled to HPLC is more common, and the charge-state logic above applies to both.

What are the common mistakes when reading ESI spectra of peptides?

The most common mistakes are assuming the highest peak is the molecular ion, ignoring adducts, and misassigning charge when peaks are not resolved. Each has a simple remedy.

  1. Reading m/z as mass. A 2+ ion appears at about half the neutral mass, so a peak at 651.8 does not mean a peptide of 651.8 Da.
  2. Overlooking adducts. Sodium and potassium adducts appear at +21.98 and +37.96 Da relative to the protonated ion for singly charged species, and they distort the pattern.
  3. Using the wrong mass definition. Calculated and observed values must both be monoisotopic or both average, which matters because the two differ by a larger amount as peptides grow.
  4. Neglecting in-source fragments. Fragile modifications can fragment before detection, so an unexpected peak does not always mean an impurity [5].
  5. Treating a single spectrum as purity. Mass spectrometry confirms identity of the species detected, while HPLC addresses the proportion of each species. See how peptide testing works and peptide mass spectrometry.

Each BioAura batch is supplied with batch-specific documentation, including a certificate of analysis, with identity confirmed by mass spectrometry and purity verified by HPLC at greater than 99%.

What are the limits of this article?

The kisspeptin-10 values are calculated from the sequence rather than measured, and real spectra depend on the instrument and conditions. The review literature describes the principles, not product-specific results.

  1. Calculated values. The m/z figures are theoretical for the unmodified amidated sequence.
  2. Instrument dependence. Charge distributions vary with source design, flow rate, and solution composition [2,3].
  3. Review sources. Several citations are reviews [2,3,5], so primary data should be consulted for specific methods.
  4. No product inference. BioAura makes no claim about any use of kisspeptin-10 in people. This article describes mass spectrometry principles for laboratory research only.

Key Takeaways

  • m/z is not mass: ESI shows mass divided by charge, so the charge must be determined first.
  • Adjacent peaks give z: two neighbouring peaks in a series let you calculate the charge without prior knowledge.
  • Formula: neutral mass equals z times (m/z minus 1.00728) for protonated ions.
  • Short peptides, low charge: kisspeptin-10 has few basic sites, so it is expected mainly at 1+ and 2+.
  • Scope: this is analytical arithmetic with no claim about any use in people.

Frequently Asked Questions

What are charge states in electrospray mass spectrometry?
A charge state is the number of protons a peptide ion carries, written as [M+zH]z+. Electrospray typically produces a series of such ions for one peptide, each at a different m/z. The classic description is a sequence of peaks whose ions differ by one charge from their neighbours.

How do I calculate the mass of a peptide from m/z?
For a positive-mode protonated ion, multiply the charge by the m/z minus 1.00728 Da, the proton mass. For example, a 2+ ion at m/z 651.823 gives 2 x (651.823 - 1.00728) = 1301.631 Da, which matches the calculated monoisotopic mass of kisspeptin-10.

How can I find the charge of a peak?
Take two adjacent peaks in the series. With m1 the higher and m2 the lower m/z, z for the higher-m/z peak is (m2 - 1.00728) / (m1 - m2), rounded to the nearest integer. High-resolution data can also reveal charge from the spacing of isotope peaks.

Why do short peptides show fewer charge states?
Positive-mode charge is mainly carried on basic sites such as arginine, lysine, histidine, and the N-terminal amine. A short peptide with few such sites, like kisspeptin-10 with one arginine and an N-terminus, can accept only a few protons, so it appears mainly at low charge states.

Does mass spectrometry show peptide purity?
Not by itself. Mass spectrometry confirms the identity of the species detected, while HPLC measures the proportion of the main species. Ion response also differs between peptides, so relative peak heights in a mass spectrum are not a reliable purity figure.

How are BioAura peptides verified for research?
Each BioAura research product is supplied with batch-specific documentation: identity is confirmed by mass spectrometry and purity is verified by HPLC at greater than 99%, recorded on a certificate of analysis. BioAura products are sold strictly for laboratory research use only and are not intended for human or veterinary use.

Research & Sources

  1. Fenn JB, Mann M, Meng CK, Wong SF, Whitehouse CM. Electrospray ionization for mass spectrometry of large biomolecules. Science. 1989;246(4926):64-71. doi:10.1126/science.2675315. https://pubmed.ncbi.nlm.nih.gov/2675315/
  2. Kebarle P, Verkerk UH. Electrospray: from ions in solution to ions in the gas phase, what we know now. Mass Spectrom Rev. 2009;28(6):898-917. doi:10.1002/mas.20247. https://pubmed.ncbi.nlm.nih.gov/19551695/
  3. Cech NB, Enke CG. Practical implications of some recent studies in electrospray ionization fundamentals. Mass Spectrom Rev. 2001;20(6):362-87. doi:10.1002/mas.10008. https://pubmed.ncbi.nlm.nih.gov/11997944/
  4. Wilm M, Mann M. Analytical properties of the nanoelectrospray ion source. Anal Chem. 1996;68(1):1-8. doi:10.1021/ac9509519. https://pubmed.ncbi.nlm.nih.gov/8779426/
  5. Banerjee S, Mazumdar S. Electrospray ionization mass spectrometry: a technique to access the information beyond the molecular weight of the analyte. Int J Anal Chem.2012;2012:282574. doi:10.1155/2012/282574. https://pubmed.ncbi.nlm.nih.gov/22611397/